Calibration Products: Types, Uncertainty and How to Choose

Calibration products are the reference standards, calibrators and check devices you use to test measuring instruments against a known value. Choose them on four points: the quantity and range, an uncertainty well below the tolerance you verify (4:1 is a common minimum), a traceable certificate from an ISO/IEC 17025 accredited laboratory, and a recalibration interval based on drift history. Brand and price come after those four.
The sections below group the equipment by quantity, show how to check the uncertainty ratio, and explain traceability, intervals and receiving checks. For a tool-by-tool comparison, see our guide on how to choose calibration tools for industrial engineers.
What types of calibration equipment are there?
Equipment is usually ranked by its place in the traceability chain between national standards and the instruments on your plant floor.
| Tier | Role | Calibrated by |
|---|---|---|
| Reference standard | Highest accuracy kept in the organisation, used to calibrate other standards | An accredited external laboratory |
| Transfer standard | Carries a value between the laboratory and sites or production areas | Comparison against a reference standard |
| Working standard | Routine calibration of instruments in production or quality control | Comparison against a reference or transfer standard |
| Verification tool | Quick go or no-go checks between formal calibrations | Checked against a working standard |

Which calibration product do you need for each quantity?
| Quantity | Calibration products | Where to look |
|---|---|---|
| Voltage, current, resistance | Multifunction and loop calibrators | Loop calibrators, Fluke 754 guide |
| Pressure | Deadweight testers, pressure calibrators, digital reference gauges | Pressure calibrators, deadweight tester comparison |
| Temperature | Dry-block and bath calibrators, reference thermometers | Temperature calibrators, Fluke temperature calibrator guide |
| Mass | OIML weights with certificates, balances | OIML calibration weights, weights for digital scales |
| Force and torque | Reference force transducers, deadweight machines | Load cell calibrator guide |
| Sound | Acoustic calibrators, reference meters | Sound meter calibration guide |
Some examples from our range, with specifications taken from the listings:
- The Druck I1001 IRS1-B is a digital pressure gauge with a resonant pressure sensor, covering 750 to 1,150 mbar absolute, with precision quoted to 0.025 mbar. That range suits barometric reference work, such as the air pressure corrections some calibrators and mass work need.
- The Radwag C32.60.1 PM is a 60 kg by 1 g precision balance with automatic internal calibration. Pair it with certified check weights.
- The MadgeTech CurrentX16 logs eight channels of DC current to 30 mA, which fits recording loop current during 4 to 20 mA checks.
- The Keysight DAQ973MB offers up to 1 MS/s and 24-bit resolution for recording multi-channel calibration runs. It is a recorder, not a reference standard, and needs its own calibration.
EURAMET publishes calibration guidelines for specific quantities, which give a common procedure to follow.
How much better than the instrument must the standard be?
The test uncertainty ratio (TUR) is the tolerance of the instrument divided by the uncertainty of your calibration. A ratio of 4:1 is a widely used minimum. ASTM E74 for force uses 10:1 from primary to secondary standards, 5:1 from secondary to working standards and 4:1 from working standards to testing machines, as summarised by Morehouse.
Worked example. A pressure transmitter spans 0 to 10 bar with a tolerance of ±0.1 % of span, which is ±10 mbar. A 4:1 ratio needs a calibration uncertainty of 2.5 mbar or better. A reference specified at ±0.015 % of reading gives 1.5 mbar at 10 bar, so the TUR is 10 / 1.5 = 6.7:1 and the check passes. A reference specified at ±0.05 % of reading gives 5 mbar, a TUR of 2:1, and does not. Include resolution, stability, temperature effects and the certificate uncertainty when you total the calibration uncertainty, not only the accuracy specification.
What makes a calibration certificate traceable?
Metrological traceability is an unbroken chain of calibrations, each with a stated uncertainty, back to national or international standards. In practice, buy from a laboratory accredited to ISO/IEC 17025 whose scope covers your range, and check that the certificate shows:
- results with the uncertainty and coverage factor;
- the reference standards used and their traceability;
- the conditions during calibration and the calibration date;
- whether the values are as-found, as-left or both.
ILAC’s policy series includes P10 on metrological traceability and P14 on measurement uncertainty in calibration. The BIPM key comparison database lists national metrology institute capabilities behind international traceability claims, and the BIPM guides cover uncertainty evaluation. For help with the paperwork, see how to read a calibration certificate and calibration equipment for ISO/IEC 17025 laboratories.

How do you set the calibration interval?
Start with the manufacturer’s advice and the guidance of your accreditation body, then adjust from data. ILAC-G24 / OIML D 10:2022 describes review methods such as the staircase method, control charts, usage-time methods and black-box operational checks. The factors that move an interval are:
- drift seen in earlier calibrations of the same or similar equipment;
- severity of use, including handling, transport and environment;
- the consequence of an out-of-tolerance result on product quality or safety;
- contract, regulatory or accreditation limits on the maximum interval.
Shorten the interval when as-found values approach the tolerance, and lengthen it only after several stable results in a row.
What should you check when new equipment arrives?
- Inspect the packaging and the unit for shipping damage.
- Match model and serial number to the certificate and confirm its date and uncertainty.
- Compare the unit with an existing standard or check standard, and record the as-received readings.
- Enter it in your asset register with its next due date and assigned procedure.
Read our returns and warranty policy before you order. Defective or damaged items are covered by a 30-day money-back guarantee, and the manufacturer’s warranty applies after that.

How do you control the calibration environment?
Run calibrations inside the operating range that the equipment specifies, and log temperature, humidity and pressure. Mass work uses 20 °C as its reference temperature, and OIML R 111-1 defines conventional mass at 20 °C. Keep equipment out of draughts, let it acclimatise after moving between rooms, and handle transport cases as part of the control plan.
How can you reduce uncertainty without buying a better standard?
- Tighten temperature and humidity control in the calibration area.
- Take more repeat readings to improve the Type A estimate.
- Apply the certificate correction instead of using the nominal value.
- Allow proper thermal equilibration before reading.
- Train operators on the connection and procedure.
Better climate control for the calibration area often gives more benefit than a higher-accuracy tool.
Browse the calibration instruments category, or the full range of test and measurement instruments at IndoMultiMeter, shipped worldwide.
Frequently asked questions
What are calibration products?
They are the reference standards, calibrators, transfer devices and check tools used to test other instruments against known values, for example deadweight testers, loop calibrators, dry-block calibrators and certified weights.
What is a test uncertainty ratio, and what is enough?
It is the instrument tolerance divided by the calibration uncertainty. A ratio of 4:1 is a widely used minimum. ASTM E74 uses 10:1, 5:1 and 4:1 for successive tiers of force standards.
How often should calibration equipment be recalibrated?
There is no universal interval. Start from the manufacturer’s advice, then use the as-found history and the methods in ILAC-G24 / OIML D 10:2022 to shorten or lengthen it.
What is the difference between a reference and a working standard?
A reference standard is the highest-accuracy device in your organisation and is calibrated by an accredited external laboratory. A working standard is calibrated against it and is used for routine calibration of production instruments.
What should a calibration certificate show?
Results with uncertainty and coverage factor, the reference standards and their traceability, the conditions and date, and the accreditation scope. Without a stated uncertainty and traceability, the certificate does not support a traceable result.