Fluke Temperature Calibrators: Models, Ranges and How to Choose

Fluke temperature calibrators fall into two groups. Heat-source calibrators (dry-wells and metrology wells) give a sensor a stable, known temperature. The main models are the 9100S (35 to 375 °C), 9142 (-25 to 150 °C), 9143 (33 to 350 °C), 9144 (50 to 660 °C) and the 9170 series (-45 to 700 °C across four models). Sensor simulators such as the 712B (RTD) and 714B (thermocouple) replace the sensor with a known resistance or voltage to test transmitters and indicators. Choose by temperature range, the accuracy you need at that range, and whether you calibrate sensors or electronics.
Which Fluke dry-wells and metrology wells are available?
The table lists the published figures. Sources are the Fluke product pages for the 9142 family, the 9100S and the 9173. The 9170 series figures also draw on reseller tables. Confirm them on the current datasheet for your model and region.
| Model | Range | Display accuracy | Stability |
|---|---|---|---|
| 9102S handheld | -10 to 122 °C | ±0.25 °C | ±0.05 °C |
| 9100S handheld | 35 to 375 °C (some regional pages list a lower start) | ±0.25 °C at 50 and 100 °C; ±0.5 °C at 375 °C | ±0.07 °C at 50 °C; ±0.1 °C at 100 °C; ±0.3 °C at 375 °C |
| 9142 field metrology well | -25 to 150 °C | ±0.2 °C | ±0.01 °C |
| 9143 field metrology well | 33 to 350 °C | ±0.2 °C | ±0.02 °C at 33 and 200 °C; ±0.03 °C at 350 °C |
| 9144 field metrology well | 50 to 660 °C | ±0.35 °C at 50 and 420 °C; ±0.5 °C at 660 °C | ±0.05 °C at 660 °C |
| 9170 and 9171 | About -45 to 140 °C and -30 to 155 °C | ±0.1 °C | ±0.005 °C |
| 9172 and 9173 | About 35 to 425 °C and 50 to 700 °C | ±0.2 °C to 425 °C; ±0.25 °C from 425 to 660 °C | 9173: ±0.005 °C at 50 to 100 °C; ±0.01 °C to 425 °C; ±0.03 °C above |
For the 9142, Fluke also lists axial uniformity of ±0.05 °C at 40 mm, radial uniformity of ±0.01 °C, a 150 mm immersion depth, a 30 mm insert outer diameter and a weight of 8.16 kg. All specifications apply within a stated ambient range, 13 to 33 °C for the 9142 family.
What do display accuracy, stability and uniformity mean?
Display accuracy is how closely the well’s own readout matches the true block temperature. Stability is how much the block temperature varies over time once it settles. Axial uniformity is the temperature difference along the depth of the well, and radial uniformity is the difference between holes. For tight work, do not rely on the display. Put a calibrated reference thermometer in the well beside the sensor under test, so stability and uniformity set the uncertainty instead of the display accuracy. Our guide to dry-block versus calibration bath explains when a bath is the better heat source, and temperature calibration standards covers the reference thermometers. Fluke also publishes guidance on selecting a dry-block calibrator.

What does a calibration uncertainty look like with a 9142?
This worked example is illustrative. It calibrates a Pt100 at 150 °C in a 9142 with a reference thermometer in the same insert. The 9142 figures are the Fluke specifications above. The reference uncertainty of ±0.03 °C (k=2) and the 0.01 °C resolution of the device under test are assumptions. Specification limits are treated as rectangular distributions, so the standard uncertainty is the limit divided by the square root of 3. Beamex describes these uncertainty components in dry-block calibration in more detail.
| Component | Limit | Standard uncertainty (°C) |
|---|---|---|
| Reference thermometer calibration (assumed, k=2) | ±0.03 | 0.0150 |
| Well stability | ±0.01 | 0.0058 |
| Axial uniformity at 40 mm | ±0.05 | 0.0289 |
| Radial uniformity | ±0.01 | 0.0058 |
| Resolution of device under test (0.01 °C) | ±0.005 | 0.0029 |
| Combined (root sum of squares) | 0.0337 | |
| Expanded, k=2 | about ±0.07 °C |
Axial uniformity dominates, about three quarters of the variance. That is why immersion depth and sensor length matter as much as the well model. IEC 60751 tolerances at 150 °C are ±(0.30 + 0.005|t|) = ±1.05 °C for class B and ±(0.15 + 0.002|t|) = ±0.45 °C for class A. An expanded uncertainty of ±0.07 °C gives a test uncertainty ratio of about 15:1 against class B and about 6.4:1 against class A.

What are the Fluke 712B and 714B for?
These are sensor simulators, not heat sources. The 712B measures and simulates 13 RTD types and resistance. The 714B measures and simulates 17 thermocouple types and millivolts. Use them to check temperature transmitters, indicators, controllers and PLC input cards by injecting a known resistance or voltage in place of the sensor. They do not check the sensor itself. The Fluke 754 documenting process calibrator also covers temperature signals, with HART support. See our Fluke 754 guide, and the Fluke 712B and Fluke 714B product pages.
Which Fluke temperature calibrator should you choose?
| Task | Choose | Why |
|---|---|---|
| Check sensors on site, 35 to 375 °C, light kit | 9100S handheld | Portable, moderate accuracy |
| Field calibration of PRTs and thermometers, -25 to 150 °C | 9142 | Stability of ±0.01 °C and sub-zero range |
| Calibration up to 350 °C | 9143 | Moderate range with ±0.02 to ±0.03 °C stability |
| High-temperature sensors up to 660 °C | 9144 | Reaches 660 °C in a field instrument |
| Tighter display accuracy and a reference input | 9170 series | ±0.1 °C display accuracy on the cold models; reference-thermometer versions |
| Test a transmitter or indicator electronics | 712B, 714B or 754 | Simulate the sensor signal |
To calibrate a complete temperature loop, use both a well for the sensor and a simulator for the electronics.

How do you keep the calibrator traceable?
The calibrator needs its own calibration. Ask for a certificate that lists the temperature points and the uncertainty. NIST explains metrological traceability, and its GMP 11 guidance on calibration intervals describes how to set the interval from drift data. Our guide on how often instruments should be calibrated applies the same idea. Thermocouples sent for calibration can use NIST thermocouple calibration services.
Which Fluke temperature calibrators does IndoMultiMeter list?
See the Fluke 9142-C-156, the Fluke 9143-A-156, the Fluke 9100S-A-156 and the Fluke 9142-DW-156 dry-well and the full range of temperature calibrators. For other reference standards, see the secondary standard PRT guide. Send us your temperature range, insert size and certificate needs and we will confirm the model. Manufacturer warranty applies as described in our returns and warranty policy. Browse the full catalogue. If you also calibrate pressure on the same site, see the Fluke 750PA4EX intrinsically safe absolute pressure module. More about IndoMultiMeter.
Frequently asked questions
What is a Fluke temperature calibrator?
It is either a heat source, such as a dry-well or metrology well that holds a known temperature for a sensor, or a sensor simulator, such as the 712B or 714B, that sends a known resistance or voltage to test transmitters and indicators.
What is the difference between the Fluke 9142, 9143 and 9144?
They cover different ranges: the 9142 runs from -25 to 150 °C, the 9143 from 33 to 350 °C and the 9144 from 50 to 660 °C. Stability is ±0.01 °C on the 9142 and loosens at higher temperatures.
Is the display accuracy the calibration uncertainty?
No. Display accuracy describes the well readout. For tight calibration, use a calibrated reference thermometer in the well. Stability, uniformity and the reference then set the uncertainty.
Can a Fluke 712B calibrate a thermocouple or Pt100 sensor?
No. The 712B simulates and measures RTD signals to test electronics. To calibrate the sensor itself, place it in a dry-well or bath with a reference thermometer.
How often should a Fluke dry-well be calibrated?
Many programmes start at 12 months and adjust using drift history. Use the manufacturer recommendation as the starting point and review the interval with as-found data.