Test & Measurement Blog · 26 Sep 2026

Load Cell Calibrator Guide: Methods, Standards and Uncertainty

Load Cell Calibrator: Technical Guide for 2026

A load cell calibrator is a force standard used to check a load cell: either dead weights that make force from mass and local gravity, or a reference force transducer loaded in series with the cell under test. Calibration applies increasing and decreasing forces, compares the load cell output with the standard at each step, and reports error, hysteresis and repeatability with an uncertainty. Reference transducers are classified to ISO 376, force-measuring instruments are calibrated to ASTM E74, and load cells for legal-for-trade weighing are classed to OIML R 60.

Below are the calibrator types, three worked examples, the calibration procedure and the standards. For the mass standards used in deadweight work, see our guides to OIML weight classes and calibration weights for digital scales.

What types of load cell calibrator are there?

Type How force is made Best for Watch for
Deadweight machine Calibrated masses under gravity Primary standard, highest accuracy at lower forces Local gravity, air buoyancy, mass certificates
Reference transducer in a loading frame A hydraulic or screw frame loads the reference and the cell under test in series Wide force range, routine working calibrations Alignment, reference class, temperature
Lever or hydraulic amplification Dead weights multiplied by a lever or hydraulic ratio High capacities The multiplication ratio needs its own uncertainty

ASTM E74 defines primary force standards as deadweight forces applied directly, with no levers or hydraulic multipliers, using masses determined by comparison with traceable references. A summary by Morehouse gives the primary mass tolerance as 0.005 % of value, with secondary standards calibrated by comparison against primary ones. Working standards are calibrated against primary or secondary standards.

Load cell calibration traceability chain

How do you work out the force from dead weights?

Force is F = m × g × (1 − ρair / ρmass), where g is the local gravitational acceleration and the bracket is the air buoyancy correction.

Worked example 1. A 100 kg mass of density 8,000 kg/m³ sits in air of density 1.2 kg/m³, where local g is 9.781 m/s² (a typical equatorial value, so use your own measured value). Buoyancy factor: 1 − 1.2 / 8,000 = 0.99985. Force: 100 × 9.781 × 0.99985 = 977.95 N. Buoyancy alone removes 0.015 % (0.147 N). Using standard gravity, 9.80665 m/s², with no correction would give 980.67 N, which is 0.28 % too high. A deadweight machine is only as good as the local g and the mass certificates behind it.

Which standards apply?

Standard Covers Key points
ISO 376:2011 Calibration and classification of force-proving instruments Classes 00, 0.5, 1 and 2. The maximum force is applied three times before calibration. Reproducibility is measured at three mounting positions rotated by 120 degrees. A calibration is valid for a maximum of 26 months.
ASTM E74 Calibration and verification of force-measuring instruments Primary, secondary and working standards, with classes AA and A. Since E74-18 the instrument is calibrated under decreasing as well as increasing force, and the user determines uncertainty in service.
ISO 7500-1 Verification of tension and compression testing machines A class 0.5 machine must be calibrated with a class 0.5 transducer classified to ISO 376. A calibration interval of no more than 12 months is recommended.
OIML R 60-1:2021 Load cells for legal metrology Accuracy classes A to D, a limit on verification intervals and error limits that include hysteresis.
DKD-R 3-3 Simplified force calibration procedures Does not meet the requirements for calibrating materials testing machines.

The ISO 376, ISO 7500-1 and DKD-R 3-3 facts above follow a GTM overview of force standards. The ASTM E74-18 changes follow Morehouse. Buy the standards for compliance work. The PTB guideline is available as DKD-R 3-3.

What is the calibration procedure?

  1. Mount the load cell in line with the force axis, connect the indicator and let the electronics warm up as the datasheet states. Let the cell and the standard reach the same temperature.
  2. Preload by applying the maximum force three times, returning to zero between cycles.
  3. Record the zero reading.
  4. Apply increasing force in steps, using the same dwell time at each step before reading.
  5. Apply decreasing force through the same steps to expose hysteresis.
  6. Rotate the transducer and repeat. ISO 376 uses three mounting positions 120 degrees apart, and ASTM E74-18 uses roughly 120 and 240 degrees for compression.
  7. Record the zero return, then evaluate error, repeatability, hysteresis and creep, fit the calibration curve and state the uncertainty.

Worked example 2. A 50 kN load cell is rated at 2 mV/V, which is 20 mV at 10 V excitation. At 50 kN the indicator reads 2.0012 mV/V. The span error is (2.0012 − 2.0000) / 2.0000 = +0.06 % of full scale. At 25 kN the increasing reading is 1.0004 mV/V and the decreasing reading is 1.0010 mV/V. The difference is 0.0006 mV/V, or 0.03 % of full-scale output, and that is the hysteresis at that point.

For deviations from a straight line, fit a low-order polynomial to the calibration data, such as F = a0 + a1V + a2V², and check the residuals. Use the fit only inside the force range that was calibrated.

Load cell uncertainty budget components

What accuracy classes does OIML R 60 define?

OIML R 60-1:2021 ranks load cells into classes A, B, C and D by the maximum number of load cell verification intervals, nLC.

Class Lower limit of nLC Upper limit of nLC Error steps (in units of v)
A 50,000 Unlimited 0.5 up to 50,000 v, 1.0 up to 200,000 v, 1.5 above
B 5,000 100,000 0.5 up to 5,000 v, 1.0 up to 20,000 v, 1.5 up to 100,000 v
C 500 10,000 0.5 up to 500 v, 1.0 up to 2,000 v, 1.5 up to 10,000 v
D 100 1,000 0.5 up to 50 v, 1.0 up to 200 v, 1.5 up to 1,000 v

The error limit is pLC times 0.5 v, 1.0 v or 1.5 v, where v is the load cell verification interval and pLC lies between 0.3 and 0.8 (0.7 is assumed when the certificate does not state it). The limits apply to increasing and decreasing loads, so they include hysteresis. Repeatability is judged over five identical load applications for classes A and B and three for classes C and D. For creep, the change between the reading at maximum load and the reading within 30 minutes at 90 % to 100 % of Emax must not exceed 0.7 times the error limit. Source: OIML R 60-1:2021.

Worked example 3. A class C cell with nLC = 3,000 and Emax = 1,000 kg has v = 1,000 / 3,000 = 0.333 kg. With pLC = 0.7 the limit is ±0.117 kg up to 500 v (167 kg), ±0.233 kg from there to 2,000 v (667 kg) and ±0.350 kg to full scale. At the top limit, a 4:1 ratio needs a reference good to 0.0875 kg, which is 0.0088 % of 1 t.

How do you choose a load cell calibrator?

  • Capacity. The standard must reach the maximum force of the cell under test, with extra reach if you want to test overload.
  • Accuracy ratio. ASTM E74 test ratios, as summarised by Morehouse, are 10:1 from primary to secondary, 5:1 from secondary to working and 4:1 from working standards to testing machines. Treat 4:1 as a floor when you calibrate a load cell.
  • Class. For machine verification, use a transducer at least as good as the machine class, and classified to ISO 376.
  • Portability. Portable reference transducers let you check installed systems on site. A fixed deadweight or frame installation gives better environmental control.
  • Traceability. Ask for a certificate from an ISO/IEC 17025 accredited laboratory, with the uncertainty and the class stated.

Which uncertainty contributions matter most?

  • Uncertainty of the reference standard, taken from its certificate.
  • Repeatability from repeated loadings and rotations.
  • Resolution and calibration of the indicator or amplifier.
  • Misalignment, eccentric loading and side forces.
  • Temperature effects on the cell and the standard, and creep.
  • Curve fitting between calibrated points.
  • Local gravity and air buoyancy, for dead weights.

The NPL Guide to the Measurement of Force and DKD-R 3-3 cover force uncertainty in detail. ASTM E74-18 also states that determining the uncertainty of the instrument in service is the user’s responsibility, so keep the full calibration data.

How often should a load cell be calibrated?

Intervals come from your standard, regulator or quality system. ISO 376 certificates are valid for a maximum of 26 months, ISO 7500-1 recommends no more than 12 months for testing machines, and recalibration is needed after overload or repair. For everything else, use the methods in ILAC-G24 / OIML D 10:2022: review the as-found data and shorten or lengthen the interval from the drift history.

Automated calibration workflow

What about load cells in trade weighing?

When a load cell is part of a legal-for-trade scale, the scale is verified to OIML R 76-1 and the cell to R 60. The Measurement Canada evaluation manual shows how laboratory and field evaluation of load cells for trade is organised in one jurisdiction. Check the rules of the country where the scale is used.

Which equipment can we supply?

  • Ametek Chatillon DMG-500: mechanical force gauge, 500 lbf in 2.5 lbf increments, 0.5 % of full scale, supplied with a NIST certificate of calibration, as listed.
  • Ametek MG TQD-40000N/C: deadweight tester for torque calibration, supplied with certification for traceability.
  • IET Labs SCA-1NF: signal conditioning amplifier for strain gauge and bridge measurements.
  • Torbal BAH30: washdown scale with an IP68 load cell, 60 lb x 0.02 lb, which you can verify with test weights.

To record bridge output, the Keysight DAQ973MB is listed with 24-bit resolution and USB, LAN and GPIB interfaces. It needs a stable excitation source and its own calibration. Browse the torque sensors and testers category, or the full range at IndoMultiMeter, shipped worldwide.

Frequently asked questions

What is a load cell calibrator?

It is a force standard used to calibrate a load cell, either dead weights acting under gravity or a reference force transducer in a loading frame. The load cell output is compared with the standard at each force step.

How many times should a load cell be preloaded?

ISO 376 applies the maximum force three times before the calibration forces are applied. Return to zero between cycles and record the zero reading before the first step.

Should I use dead weights or a reference transducer?

Dead weights give a primary force with the lowest uncertainty at lower forces, but need local gravity and mass certificates. A reference transducer in a loading frame covers wider ranges and is faster for routine work, provided its class and certificate suit the cell under test.

How often should a load cell be calibrated?

There is no single interval. ISO 376 certificates are valid for a maximum of 26 months and ISO 7500-1 recommends no more than 12 months for testing machines. Review the as-found drift, and recalibrate after overload or repair.

What does C3 mean on a load cell?

C3 means accuracy class C with up to 3,000 load cell verification intervals. For a 1,000 kg cell, the interval v is 0.333 kg, and the OIML R 60 error limit is pLC times 0.5 v up to 500 v.