Calibration Weights for Digital Scales: Classes and Selection

Choose calibration weights by class: under OIML R 111-1 the weight’s error must be no more than one third of the scale’s error limit at the test load. For an industrial scale that is usually class M1 or F2, and for a 0.1 mg analytical balance it is usually E2 or E1. Add a certificate when the result must be traceable, and handle the weights with gloves or forceps so their mass does not change.
This guide explains the classes, tabulates the error limits from OIML R 111-1, works through a scale example, and covers certificates, buoyancy, handling and intervals. For a closer look at the tolerances, see OIML weight classes and tolerances.
What do the OIML weight classes mean?
OIML R 111-1:2004 defines nine classes: E1, E2, F1, F2, M1, M1-2, M2, M2-3 and M3. Each class is tied to what the weight is used for.
| Class | Intended use in R 111-1 |
|---|---|
| E1 | Traceability between national mass standards and E2 weights and lower. Always supplied with a calibration certificate. |
| E2 | Verification or calibration of F1 weights, and use with special accuracy class I instruments. Always supplied with a certificate. |
| F1 | Verification or calibration of F2 weights, and use with class I and high accuracy class II instruments. |
| F2 | Calibration of M1 and possibly M2 weights, and important commercial transactions on high accuracy class II instruments. |
| M1 | Calibration of M2 weights, and use with medium accuracy class III instruments. |
| M2 | Calibration of M3 weights, general commercial transactions and class III instruments. |
| M3 | Class III and ordinary accuracy class IIII instruments. |
M1-2 and M2-3 are lower-accuracy classes for weights from 50 kg to 5,000 kg. The instrument classes (I, II, III, IIII) come from OIML R 76-1 for non-automatic weighing instruments.

What are the maximum permissible errors?
The table lists the maximum permissible error (±, in mg) from Table 1 of OIML R 111-1:2004. The limits relate to conventional mass and must not be extrapolated beyond the nominal values the recommendation gives for each class.
| Nominal value | E1 | E2 | F1 | F2 | M1 | M2 | M3 |
|---|---|---|---|---|---|---|---|
| 20 kg | 10 | 30 | 100 | 300 | 1,000 | 3,000 | 10,000 |
| 10 kg | 5.0 | 16 | 50 | 160 | 500 | 1,600 | 5,000 |
| 5 kg | 2.5 | 8.0 | 25 | 80 | 250 | 800 | 2,500 |
| 2 kg | 1.0 | 3.0 | 10 | 30 | 100 | 300 | 1,000 |
| 1 kg | 0.5 | 1.6 | 5.0 | 16 | 50 | 160 | 500 |
| 500 g | 0.25 | 0.8 | 2.5 | 8.0 | 25 | 80 | 250 |
| 200 g | 0.10 | 0.3 | 1.0 | 3.0 | 10 | 30 | 100 |
| 100 g | 0.05 | 0.16 | 0.5 | 1.6 | 5.0 | 16 | 50 |
The limit is an allowance on the error, not a measured error. A 1 kg class F1 weight may deviate from nominal by up to 5.0 mg and still conform. In addition, the expanded uncertainty (k = 2) of its conventional mass must be no more than one third of the limit.
How do you choose a class for a digital scale?
Work from the error limit of the scale. OIML R 111-1 notes that the error of a weight used to verify an instrument shall not exceed one third of the instrument’s maximum permissible error, and OIML R 76-1 repeats the rule in clause 3.7.1. For class E2 or better, the weight’s uncertainty may be used instead of its error, provided its actual conventional mass is taken into account.
Worked example 1, a class III scale. The scale has a capacity of 15 kg and a verification interval e = 5 g, so 3,000 e. R 76-1 Table 6 gives class III an initial verification limit of ±1.0 e for loads from 500 e to 2,000 e. At a 10 kg test load (2,000 e) the limit is ±5 g. One third of that is 1.67 g, or 1,667 mg. From the table, a 10 kg weight has a limit of 500 mg for M1, 1,600 mg for M2 and 5,000 mg for M3. M1 passes with room to spare, M2 passes only barely and M3 fails. Use M1, or F2 if you want more margin.
Worked example 2, a 0.1 mg balance. You verify a 200 g load against a tolerance of ±1.0 mg that you set yourself. One third of that is 0.33 mg. At 200 g, E2 allows 0.3 mg and E1 allows 0.10 mg, so both pass. F1 allows 1.0 mg and uses up the whole tolerance on its own.
For large scales, R 76-1 permits another constant load to replace part of the standard weights at the place of use, provided standard weights of at least half of Max are used.
What does a digital scale test look like?
The minimum test scheme in EURAMET Calibration Guide No. 18 is a good template for non-automatic scales.
- Level the scale, allow warm-up as the manufacturer states, and let the weights reach room temperature.
- Zero the scale with an empty pan.
- Check errors of indication at five or more loads spread across the range, for example zero or the minimum load, 0.25 Max, 0.5 Max, 0.75 Max and Max.
- Check repeatability by applying one load at least five times (three times if the load is 100 kg or more).
- Check eccentricity with a load of about Max/3 or more at five positions: centre and four off-centre.
- Record the readings, the weight identification and class, temperature and operator, and compare each error with the tolerance.

Do you need a certificate with the weights?
E1 and E2 weights always come with a calibration certificate under R 111-1, and the user must take the certificate deviation from nominal into account. For other classes the certificate is a separate choice, and you need one when the result must be traceable under ISO/IEC 17025 requirements. Class and certificate vary independently. The listings for the Troemner 7408-F1 (class F1, no certificate) and the Troemner 7428-F1W (class F1, NVLAP accredited certificate) show the difference.
NIST’s mass calibration program shows how national metrology institutes anchor the traceability chain. A2LA lists common certificate findings, such as missing uncertainty statements or traceability declarations. A useful certificate states the weight identification, the conventional mass value, the expanded uncertainty and coverage factor, the conditions, the reference standards and the accreditation scope. For help reading one, see how to read a calibration certificate. ILAC’s policy series covers uncertainty and traceability in calibration.

Does air buoyancy matter?
R 111-1 defines conventional mass as the mass of a reference weight of density 8,000 kg/m³ that balances the weight in air of density 1.2 kg/m³ at 20 °C. A 1 kg weight of density 8,000 kg/m³ has a volume of 125 cm³, so at 1.2 kg/m³ the air it displaces weighs 150 mg. That is why the standard ties conventional mass to reference conditions.
Altitude lowers air density. The recommendation says class E weights used above 330 m must come with their density and its uncertainty, and so must class F1 weights above 800 m. If you work at altitude with E or F1 weights, ask for that data.
How should you handle and store calibration weights?
- Use lint-free gloves, forceps or a handling tool. Skin oils and dust change mass.
- Lift by the knob or handle and place weights gently, centred on the pan.
- Keep sets in their fitted case with each weight in its own compartment, and keep the case closed.
- Do not use calibration weights for anything else, and record any drop or knock.
- Clean only as the maker advises, and never use abrasives.
Construction varies by class. R 111-1 requires class E and F weights to have non-porous, glossy surfaces. Class M cast weights from 100 g to 50 kg have a finish like grey cast iron cast in a fine sand mould, so they need protection against rust in damp stores. The listings for our sets show stainless steel or Troemner Alloy 8 for the precision classes. For example, the Troemner 7418-F2W is listed in Alloy 8 with a density of 8.03 g/cm³, and the Rice Lake E2 set as polished stainless steel with a density of 7.95 g/cm³.
How often should weights be recalibrated?
There is no single interval. R 111-1 leaves subsequent verification limits to each state, and says that if the allowed error is larger than the Table 1 limit, the weight can no longer be declared to belong to that class. Use the methods in ILAC-G24 / OIML D 10:2022: record each as-found value, compare it with the limit, shorten the interval if weights drift, and lengthen it only after several stable results. Use a higher-class weight as a check standard between calibrations.
Which weight sets can we supply?
- Troemner 7408-F1: 17 class F1 weights, 10 kg to 1 g, listed without a certificate.
- Troemner 7428-F1W: 21 class F1 weights, 100 g to 1 mg, with an NVLAP accredited certificate.
- Troemner 7418-F2W: 12 class F2 weights, 500 g to 1 g, with an NVLAP accredited certificate.
- Rice Lake class E2 set: five weights, 10 kg to 1 kg, with an accredited certificate per OIML R 111, as listed.
These details come from the product listings. Confirm class, range and certificate on the listing before you order. Browse all OIML calibration weights, or the analytical balance calibration guide if you work at milligram level. The full range of instruments is at IndoMultiMeter, shipped worldwide.
Frequently asked questions
Which class of calibration weights do I need for my scale?
Take the scale’s error limit at the test load and divide it by three. Pick the class whose limit at your test weight is at or below that value. For a class III scale with 5 g verification intervals tested at 10 kg, the limit is 5 g, so a weight limit of 1.67 g or less works, and M1 (500 mg) passes.
Do calibration weights need a certificate?
E1 and E2 weights always come with one under OIML R 111-1. For other classes it depends on whether you need traceable results, for example under ISO/IEC 17025. Check that the certificate states the conventional mass, the uncertainty and the traceability.
What is the tolerance of a 1 kg M1 weight?
The maximum permissible error of a 1 kg class M1 weight is ±50 mg under OIML R 111-1:2004. For comparison, 1 kg limits are ±16 mg for F2, ±5.0 mg for F1 and ±1.6 mg for E2.
How should I handle and store calibration weights?
Wear lint-free gloves or use forceps or a handling tool, lift by the knob, store each weight in its own compartment of the case, and record any drop. Fingerprints and dust change the mass.
How often do calibration weights need recalibrating?
There is no fixed interval. Set it from the as-found history: shorten it if weights drift and lengthen it after several stable results, using a higher-class check standard between calibrations.