Analytical Weighing Balance Calibration Guide

For analytical balance calibration, load the balance with traceable reference weights at five or more points from zero to maximum capacity, repeat one load several times to measure repeatability, move a load to five pan positions to check eccentricity, and report every error with its measurement uncertainty. The test scheme in EURAMET Calibration Guide No. 18 is the common reference for this. Between calibrations, a short repeatability check shows whether the balance is still fit for the smallest sample you weigh.
For a deeper look at weight grades, see our guide to OIML weight classes and tolerances. If you are still choosing a balance, read analytical vs precision balance.
What tests make up an analytical balance calibration?
EURAMET cg-18 (version 4.0, 2015) specifies three tests for non-automatic weighing instruments. The table summarises what the guide asks for.
| Test | What it shows | Minimum scheme in EURAMET cg-18 |
|---|---|---|
| Repeatability | Scatter when the same load is placed again and again | At least one load, applied at least 5 times (3 times if the load is 100 kg or more). A load of about 0.5 Max to Max is quite common. |
| Errors of indication | Accuracy over the weighing range | At least 5 loads spread evenly, for example zero or Min, 0.25 Max, 0.5 Max, 0.75 Max and Max. |
| Eccentricity | Effect of placing the load off centre | Test load of about Max/3 or higher at five positions: centre, front left, back left, back right, front right. |
The guide also recommends recording the air temperature at least once during the calibration, and letting standard weights acclimatise to the room so that temperature differences do not shift their apparent mass.

How do you prepare the balance and the room?
- Level the balance with its spirit level and feet, and keep it away from draughts, vents, windows and vibration.
- Let the balance warm up as the manufacturer states, and let the weights reach room temperature before use.
- Handle weights with gloves or tongs, never bare hands, and inspect them for dust or marks.
- Record temperature, humidity and air pressure, because they feed the buoyancy correction and the uncertainty budget.
Humidity matters for static charge. Mettler Toledo’s guide Weighing the Right Way says relative humidity should ideally sit between 45 and 60 %, and it warns against plastic weighing vessels when humidity falls below roughly 30 to 40 %, where insulating materials can charge up and disturb the reading.
Which class of calibration weights should you use?
Weights are graded by OIML R 111-1. The table lists the maximum permissible error (MPE, in mg) for the classes used with analytical balances, copied from Table 1 of OIML R 111-1:2004.
| Nominal value | Class E1 | Class E2 | Class F1 |
|---|---|---|---|
| 1 g | ±0.010 mg | ±0.03 mg | ±0.10 mg |
| 10 g | ±0.020 mg | ±0.06 mg | ±0.20 mg |
| 50 g | ±0.03 mg | ±0.10 mg | ±0.3 mg |
| 100 g | ±0.05 mg | ±0.16 mg | ±0.5 mg |
| 200 g | ±0.10 mg | ±0.30 mg | ±1.0 mg |
Two rules from the recommendation decide how good a set you need. The weight’s expanded uncertainty (k = 2) must be no more than one third of its MPE. And for class E1 and E2 weights, which always come with a certificate, the user must take the certificate deviation from nominal into account. Pick a class whose MPE is small against the tolerance you are verifying. A 0.1 mg balance weighed against a 200 g F1 weight (±1.0 mg) has no margin left, while an E1 weight (±0.10 mg) does.
How do you work out the minimum weight of a balance?
The minimum weight is the smallest sample you can weigh while keeping the balance error inside your limit. For pharmaceutical work, USP general chapter <41> sets the repeatability test: weigh one test weight at least 10 times, and the result passes if 2 times the standard deviation, divided by the smallest net weight you intend to weigh, is 0.10 % or less. A&D’s explanation of the rule, What is minimum weight, states it as:
Minimum weight = 2000 x standard deviation
If the measured standard deviation is smaller than 0.41 d (d = scale interval), 0.41 d is used instead. Worked example for a 0.1 mg balance:
| Measured standard deviation (10 weighings) | Value used | Minimum weight |
|---|---|---|
| 0.10 mg | 0.10 mg | 200 mg |
| 0.03 mg (below 0.41 d = 0.041 mg) | 0.041 mg | 82 mg |
Repeatability depends on the day: draughts, vibration and operator skill all change it. Weigh samples comfortably above the minimum, not just at it. For the accuracy side, the same A&D summary states that a test weight of 5 % to 100 % of capacity should stay within 0.10 % of its value, and that the weight’s MPE or calibration uncertainty should be no more than a third of that limit.

What goes into the measurement uncertainty?
The uncertainty budget combines the standard uncertainties of each contribution by root sum of squares. Expanded uncertainty is the combined value times a coverage factor. EURAMET cg-18 says k is chosen for a coverage probability of 95.45 %, which is k = 2 for a normal distribution with sufficient degrees of freedom. Typical contributions are:
- repeatability of the balance, from the repeatability test;
- resolution of the display, which is half a scale interval either side of the reading;
- uncertainty of the reference weights, from their certificate;
- eccentricity, from the five-position test;
- air buoyancy and temperature effects, mainly for the most accurate work.
For routine laboratory balances, the certificate result is used in two ways: to confirm the errors stay inside your tolerance, and to feed the minimum weight and weighing-uncertainty estimates.
What should the calibration certificate show?
- Balance identification: model, serial number, location.
- Conditions during the calibration: temperature, humidity, air pressure.
- The reference weights used: identification, class and uncertainty, with traceability to national standards.
- As-found and as-left errors at every test point, plus repeatability and eccentricity results.
- Expanded uncertainty with the coverage factor stated.
- Any limits on the result, for example the load range covered.
What checks should you run between calibrations?
The 2013 revision of the USP balance chapters moved from fixed daily checks to a risk-based approach, as Pharmaceutical Technology notes. The article calls for regular calibration with uncertainty, plus periodic sensitivity and repeatability testing, and leaves check type and frequency to a risk assessment. A practical routine:
- Before use: check level, zero and draught shield, and weigh one check weight near your typical load.
- Periodically: repeat the 10-weighing repeatability test with a small weight, and confirm the minimum weight still holds.
- Trend the results on a control chart so drift shows before an error exceeds tolerance.
- Set the external calibration interval by risk: criticality of the measurement, history of the as-found errors, usage intensity and the manufacturer’s advice.
What are the common causes of a drifting or unstable balance?
- Draughts and vibration: move the balance, use the draught shield, add an anti-vibration table.
- Static charge: typical in dry rooms with plastic or glass vessels. Use metal or anti-static vessels and an ionizer.
- Temperature differences: let samples and weights acclimatise before weighing.
- Dirty or damaged weights: a fingerprint changes the mass. Clean, store in the case and send weights for recalibration on schedule.
- Linearity or eccentricity errors beyond specification: usually mechanical or electronic, so the balance needs service.
Browse analytical balances and OIML calibration weights. For other instruments in a laboratory calibration programme, browse the catalogue at IndoMultiMeter.
Looking for this instrument? IndoMultiMeter supplies Adam Equipment SAB 224e Solis Analytical Balance, 220g x 0.0001g and Sartorius QUINTIX 35-1S Analytical Balance, 30 g x 0.00001 g. Browse our full range of test & measurement instruments, shipped worldwide.
Frequently asked questions
How often should an analytical balance be calibrated?
There is no single interval. Set it from risk: how critical the measurement is, how much the as-found errors drift between calibrations, how heavily the balance is used and what the manufacturer advises. Shorten the interval if the as-found data show drift, and lengthen it only when several results in a row are stable.
What is the minimum weight of an analytical balance?
It is the smallest net sample for which the balance error stays within your limit. Under the USP repeatability rule it is 2000 times the standard deviation of at least 10 weighings, with 0.41 times the scale interval used if the measured deviation is smaller. A 0.1 mg balance with a 0.10 mg standard deviation therefore has a minimum weight of 200 mg.
Which weight class do I need for a 0.1 mg balance?
Choose the class from the tolerance you must verify. The MPE of the weight should be small against that tolerance, and OIML R 111-1 requires the weight’s expanded uncertainty to be at most one third of its MPE. At 200 g, class E1 allows ±0.10 mg, E2 ±0.30 mg and F1 ±1.0 mg.
Why test eccentricity?
Samples are rarely placed exactly at the centre of the pan. EURAMET cg-18 places a load of about one third of maximum or more at the centre and four off-centre positions, so you see how much the reading changes with position.
Does humidity affect balance calibration?
Yes, mainly through static charge and moisture uptake. Mettler Toledo suggests 45 to 60 % relative humidity as the ideal range, and advises against plastic weighing vessels below about 30 to 40 %. Record humidity and temperature during every calibration.