Tester Calibration: Traceability, TUR and Decision Rules

Tester Calibration: Standards and Best Practices

Calibrating a tester, meaning any test instrument such as a multimeter, pressure gauge, loop calibrator or torque tester, means comparing it with a traceable reference at several points, recording the errors with an uncertainty, and judging them against a tolerance using a stated decision rule. The reference should be clearly better than the instrument, usually with a test uncertainty ratio (TUR) of at least 4:1. Then set the next interval from the as-found results. Without traceability, an uncertainty and a decision rule, a check is only a verification.

What does metrological traceability mean?

Traceability is the property of a result that lets it be related to a reference through a documented, unbroken chain of calibrations, each adding to the measurement uncertainty. NIST’s description of metrological traceability uses this definition. In practice the chain looks like this:

Level Example What you must hold
National or international standard National metrology institute Reference to the SI
Accredited calibration laboratory ISO/IEC 17025 laboratory Certificate with uncertainty for the lab’s reference
Your reference standard Calibrator, pressure standard, deadweight tester Valid certificate, with drift history
Your working tester Gauge, meter, torque tester Calibration record with as-found and as-left data

Metrological traceability chain

Each link needs a certificate that states the uncertainty. Read how to read a calibration certificate before you accept one, and calibration equipment for ISO/IEC 17025 laboratories for lab requirements.

How do you calibrate a tester, step by step?

  1. Identify the instrument, its tolerance (from the datasheet or your process) and the points to test, spread over the working range.
  2. Select a reference with a valid certificate and an uncertainty well below the tolerance.
  3. Stabilise the instrument and the reference in the lab conditions, and record temperature and humidity.
  4. Record as-found errors at each point, up and down for devices with hysteresis.
  5. Adjust only if an error is outside the limit, then record the as-left errors.
  6. Calculate the uncertainty, compare the errors with the tolerance using the decision rule, and issue the record or certificate.

For one specific example, see our step-by-step pressure gauge calibration procedure. The terms are explained in calibration, verification and adjustment.

Worked example: uncertainty and TUR for a pressure gauge

The numbers are illustrative. A 0 to 10 bar gauge has a tolerance of ±0.05 bar. Reference certificate: U = 0.0075 bar at k = 2. Gauge resolution is 0.02 bar. Repeatability of the readings is 0.004 bar (standard uncertainty).

Contribution Standard uncertainty (bar)
Reference (0.0075 / 2) 0.00375
Gauge resolution (half step 0.01 / √3) 0.00577
Repeatability 0.00400
Combined (root sum of squares) 0.00796
Expanded, k = 2 0.0159

The TUR is tolerance divided by expanded uncertainty: 0.05 / 0.0159 = 3.1 to 1, which is below 4:1. This is the case where a decision rule matters. The method follows the Guide to the Expression of Uncertainty in Measurement.

Uncertainty budget components

What is a decision rule and a guard band?

A decision rule states how you decide pass or fail given the uncertainty. ISO/IEC 17025 requires an accredited laboratory to agree and document the rule when it states conformity. As summarised in Tektronix’s decision rule guide, two common rules are:

  • Simple acceptance: the acceptance limit equals the tolerance limit. It is easy, but it carries a significant risk of false acceptance near the limit.
  • Guard-banded acceptance: the acceptance limit is moved inside the tolerance by a guard band, so that the probability of false accept stays low. ANSI/NCSL Z540.3 uses a rule that targets no more than 2 % false accept, and its guidance asks for a TUR above 4:1 where attainable.

Continuing the example with a guard band equal to the expanded uncertainty: the acceptance limit is 0.05 – 0.0159 = ±0.034 bar. An as-found error of 0.040 bar passes under simple acceptance and fails under the guard-banded rule. Decide the rule before you calibrate, and write it on the record.

Guard band decision rules

How often should a tester be calibrated?

There is no universal interval. ILAC-G24 / OIML D 10 describes methods for setting and reviewing it, including a simple staircase rule: lengthen the interval when the as-found error is within, for example, 80 % of the maximum permissible error, and shorten it when it is outside. Details are in our article on software and AI in calibration and in how often should instruments be calibrated. Calibrate sooner after a drop, an overload or a repair.

What should you record?

  • Instrument identification, location and owner.
  • Reference standards used, with certificate numbers and due dates.
  • Environmental conditions.
  • As-found and as-left data at every point, with uncertainty and coverage factor.
  • The tolerance, the decision rule and the pass or fail result.
  • Technician, date and next due date.

What safety points apply to electrical testers?

Use equipment rated for the measurement category of the circuit, and follow IEC 61010 for the instrument you are testing. Inspect leads before each use, and de-energise the circuit when a procedure allows. For multimeters, our Keysight multimeter guide covers specifications and calibration.

What tools support a calibration lab?

Torque testers are calibrated with a reference such as the Ametek MG TQD-40000N/C deadweight tester for torque calibration. For lab layout, read complete calibration lab setup. Browse the full range of test and measurement instruments for references and working testers.


See the whole catalogue at IndoMultiMeter. Orders ship worldwide.

Frequently asked questions

What is tester calibration?

It is the comparison of a test instrument with a traceable reference at several points, with the errors recorded together with an uncertainty. The result is judged against a tolerance using a decision rule, and as-found and as-left data are kept.

What TUR do I need?

A TUR of at least 4:1 is the usual target, meaning the tolerance is four times the expanded uncertainty of the calibration. Below that, use a guard-banded decision rule or a better reference.

What is a guard band?

A guard band moves the acceptance limit inside the tolerance limit by an amount related to the uncertainty, which cuts the chance of accepting an out-of-tolerance instrument. With U = 0.0159 bar and a tolerance of 0.05 bar, the acceptance limit becomes 0.034 bar.

What is the difference between as-found and as-left data?

As-found data is what the instrument read before any adjustment, and it shows how it behaved in service. As-left data is what it reads after adjustment. Keep both, because as-found trends set the interval.

How often should a tester be calibrated?

Set it from risk and history. ILAC-G24 describes methods such as the staircase rule, which lengthens the interval when the as-found error is within about 80 % of the permissible error and shortens it when the error is outside it.