Test & Measurement Blog · 11 Nov 2026

PRT Ice Point Check: How to Detect R0 Drift in 6 Steps

PRT Ice Point Check: How to Detect R0 Drift in 6 Steps

To check a secondary standard PRT at the ice point, immerse it in a properly made, stirred or well-packed ice bath of pure water, read its resistance with the same readout and excitation current used on its certificate, and compare the result with the baseline R0 on file. A shift larger than your own uncertainty limit means the probe has drifted or been disturbed, and it should be re-checked, annealed if its range allows, or recalibrated.

Why check a reference PRT at 0 °C

Secondary standard PRTs are very stable, but not immutable. Shock, vibration and thermal cycling can strain the platinum element and shift its resistance. Fluke Calibration notes that even a light tap may change the resistance at the triple point of water (Rtp) by as much as 1 mK, and that annealing at 660 °C for an hour relieves most such strains. A quick check at 0 °C catches this before the probe is used as a reference for other instruments.

The logic is the same one used in ITS-90 practice. Interpolation relies on a resistance ratio, W = R(t)/R(0.01 °C), rather than on absolute resistance, so a change in the 0 °C-region resistance is the earliest and most useful sign that the probe has moved. Fluke’s guidance on producing the highest accuracy from SPRTs says the triple-point resistance should be determined immediately after a high-accuracy measurement, and that the check can be done less often as the required uncertainty loosens. For a 1 mK requirement, Rtp is measured after each Rt measurement; for 20 mK or looser, below 420 °C and with a quite stable probe, Fluke suggests roughly once a week.

Ice point versus triple point

The ice point is a practical substitute, not an equal. ITS-90 defines the ratio against the triple point of water (0.01 °C), and ASTM E563 describes the preparation and use of an ice-point bath as a reference temperature. It covers resistance thermometry as a fallback where a triple-point cell is not available or not needed, and it provides an industrial method for relating ice-point and triple-point values.

Bath quality sets the limit of what you can resolve. Fluke specifies its 7911A2 stirred ice bath at ±0.002 °C uniformity and stability, based on a properly made ice bath mixture of pure distilled or demineralised water. Fluke also states that a stationary ice mixture can carry considerable uncertainty. If your bath is hand-made, treat apparent changes of a few millikelvin with caution until you have repeated the check.

Which PRT and what its specification supports

Use the specification of the probe to decide what a pass looks like. The figures below are taken from Fluke’s product pages. They are calibration and drift statements, not a guarantee of what your own bath will show.

Probe Range Nominal resistance (per Fluke) Calibrated accuracy at 0 °C Drift statement
Fluke 5626 −200 to 661 °C 100 Ω (±1 Ω) at RTPW ±0.006 °C (k=2), includes calibration and 100 h drift <0.006 °C per 100 h at 661 °C
Fluke 5628 −200 to 661 °C 25.5 Ω (±0.5 Ω) at RTPW ±0.006 °C (k=2), includes calibration and 100 h drift <0.004 °C per 100 h at 661 °C
Fluke 5615-12 −200 to 420 °C 100 Ω ±0.10 Ω at 0 °C ±0.012 °C See Fluke datasheet

Two practical points follow. First, the 5626 and 5628 share the same stated range, but the 25.5 Ω 5628 gives a smaller resistance change per degree, so your readout resolution matters more when you convert a resistance shift to millikelvin. Second, the 5615 is rated only to 420 °C, which limits any annealing you might consider. For model selection in depth, see our comparison of secondary standard PRTs.

Equipment you need

  • A readout or resistance bridge with a current valid calibration, able to use the same excitation current as your last certificate.
  • A dewar or insulated flask, pure distilled or demineralised water, and ice made from the same quality of water.
  • A way to crush or shave the ice finely, and a means of draining excess meltwater.
  • The probe’s last certificate, with its baseline resistance and the date and conditions of measurement.
  • A clean, dry holder or stand so the probe is not knocked while it equilibrates.

Step-by-step procedure

1. Confirm the baseline you are comparing against

Find the resistance at 0 °C or at the triple point from the most recent certificate or your own check history. Note the excitation current, the readout and the date. A comparison is only meaningful if it uses the same current and a readout of known performance, otherwise you may be measuring the readout rather than the probe.

2. Prepare the ice bath

Fill the dewar with finely crushed or shaved ice made from pure water. Add chilled pure water until the ice is just floating in slush, then drain surplus meltwater so the mixture stays compact. Follow ASTM E563 for the preparation details. Impurities in either the water or the ice are the most common reason a hand-made bath reads off 0 °C.

3. Pre-cool and immerse the probe

Hold the probe in a separate bath of ice and water for a few minutes before moving it, so it does not melt a pocket of ice around itself. Handle it gently to avoid mechanical shock. Fluke recommends at least 12.7 cm (5 in) immersion for the 5626 and 5628; allow comparable depth for any PRT, and confirm the figure for your model before you rely on it. Keep the sensing end clear of the dewar wall and bottom.

4. Let the probe settle and read

Wait until readings stop trending. Then record a series of readings at the same current as the baseline, and at a steady pace. Note the standard deviation of the series. A noisy series points to an unstable bath or a stirring problem, not to the probe.

5. Convert the shift to temperature

Subtract the baseline resistance from your reading, then divide by the probe’s sensitivity near 0 °C. For a 100 Ω PRT this is roughly 0.39 Ω per °C, so 1 mK is about 0.4 mΩ; for the 25.5 Ω 5628 it is about a quarter of that. These are rule-of-thumb figures from the standard platinum curve, so use the coefficients on your own certificate when you report a result. Expressing the shift in millikelvin lets you compare it with your bath and measurement uncertainty.

6. Decide and record

Set the action limit from your own uncertainty budget before you take the measurement, not after. A common approach is to define a warning level and a reject level as fractions of the uncertainty you claim for the probe as a reference. Then:

  • Within the warning level: log the result and keep the probe in service.
  • Between warning and reject: repeat the check with a freshly made bath. If the shift repeats, treat it as real.
  • Beyond the reject level: anneal if the probe’s range allows, then repeat the check, or send it for recalibration.

Annealing: when and how far

Fluke states that if a probe’s maximum temperature is below 660 °C, it should be annealed at its maximum temperature. For the 5626 and 5628, rated to 661 °C, annealing for about an hour at 660 °C is the described method for relieving shock strain. Never exceed a probe’s rated range: a 5615 should not be taken above 420 °C. Always repeat the 0 °C check after annealing and after the probe returns to room temperature.

Common causes of a bad result

  • Impure water or ice: contamination depresses the bath temperature.
  • Bath not compact or not stirred: warm water pockets pull the reading up.
  • Insufficient immersion: stem conduction biases the reading.
  • Different excitation current: self-heating differs from the baseline.
  • Probe knocked or moved during the series: a physical shift in resistance can appear mid-test.

If the probe passes but you still doubt the readout, a second PRT in the same bath is a simple cross-check. For the wider picture of reference thermometers and what each one is used for, see temperature calibration standards. Reference probes and other calibration instruments are listed in our catalogue, and the manufacturer’s 5626 and 5628 page carries the full specification.

Frequently asked questions

How often should a secondary PRT be checked at the ice point?

It depends on the uncertainty you claim. Fluke suggests measuring Rtp after each Rt measurement for 1 mK work, and roughly weekly for 20 mK or looser requirements below 420 °C with a stable probe. Also check after any shock, drop or unusual thermal exposure.

Can an ice bath replace a triple point of water cell?

Not for the lowest uncertainties. ITS-90 defines resistance ratios against the triple point of water (0.01 °C). ASTM E563 treats the ice point as a practical reference and fallback for resistance thermometry, with a method for relating ice-point and triple-point values.

What does a shift in R0 tell me?

It indicates the probe's resistance has changed, from strain, contamination or drift, or that the bath or readout is at fault. Repeat the check with a fresh bath first. If the shift persists beyond your action limit, anneal within the probe's rating or recalibrate.

Is an ice point check the same as calibration?

No. It verifies one point against your baseline and shows whether the probe has moved. Calibration characterises the probe over its range with a stated uncertainty and produces a certificate. The check supports your decision on when recalibration is needed.