How to Compensate an Oscilloscope Probe: Step-by-Step

How to Compensate an Oscilloscope Probe: Step-by-Step

To compensate an oscilloscope probe, connect it to the scope’s compensation terminal, display the square wave, and turn the probe’s trimmer until the waveform tops are flat with no overshoot or rounding. Do it for every probe and channel pairing, and repeat it whenever you move a probe to a different channel or scope. It takes about a minute and removes a common source of distorted edges.

What probe compensation actually does

A passive 10:1 probe forms a voltage divider with the scope input. For the divider to stay accurate across frequency, the probe’s capacitance has to be matched to the input characteristics of the channel it is connected to. Compensation is the adjustment that makes that match. Tektronix describes it as tuning the probe’s capacitance to the oscilloscope’s input impedance, and Keysight tells users to compensate a probe to match its characteristics to the oscilloscope for the best measurements.

If the match is wrong, the display no longer reproduces the signal faithfully. Tektronix lists overshoot, undershoot or signal attenuation as the symptoms of an impedance mismatch. Edge shape and pulse rise-time readings are the first to suffer, because they depend on the high-frequency end of the divider.

Compensation is not calibration

The two terms are often mixed up. Tektronix defines calibration as evaluating a probe’s measurements against a known standard, and compensation as an adjustment of the probe. Compensating a probe does not change its calibration status, and a calibration certificate does not remove the need to compensate. The wider distinction between calibration, verification and adjustment is a separate topic, and the same logic applies: an adjustment is not a calibration.

When to compensate

Tektronix recommends compensating when you connect a new probe and before taking crucial measurements. It also recommends doing it after significant environmental changes or after any maintenance on the probe. In practice, use this checklist:

  • A new probe is attached to a scope for the first time.
  • A probe is moved to another channel. Channel input capacitance is not guaranteed to be identical, so treat each pairing separately.
  • A probe is shared between different oscilloscopes.
  • You are about to measure edges, pulses or other fast signals where shape matters.
  • The probe has been dropped, repaired or fitted with different accessories.

What you need

  • The oscilloscope with a front-panel compensation terminal (often labelled Probe Comp or Demo), or an external square wave source. Keysight specifies a 1 kHz square wave source for its N28xxA/B passive probes.
  • The probe with its ground lead or ground spring and a standard probe tip.
  • The adjustment tool supplied with the probe. Tektronix advises a non-magnetic tool for the trimmer.
  • The scope and probe manuals, because trimmer location differs between probe models and some scopes compensate partly in software.

Step-by-step procedure

  1. Select the right probe setting. If the probe has a 1:1 / 10:1 switch, set it to 10:1 before compensating. Compensation is a 10:1 adjustment, and the 1:1 position bypasses the network you are trying to tune.
  2. Set the channel input. Use the 1 MΩ input. Keysight states that its N2140A probe is compatible with InfiniiVision oscilloscopes with 1 MΩ input. Match the channel’s probe attenuation setting to the probe.
  3. Connect the probe. Attach the probe to the channel, hook the tip to the compensation terminal and clip the ground lead to the adjacent ground terminal, as the Keysight InfiniiVision user’s guides describe.
  4. Display the signal. Press Autoscale. For the N28xxA/B probes, Keysight’s datasheet says to adjust the scope to show two to three cycles of the waveform over two to six vertical divisions.
  5. Adjust low-frequency compensation. Turn the probe’s LF adjustment for the flattest pulse possible. Make small turns and let the display update between each one.
  6. Adjust high-frequency compensation, if the probe has it. Keysight states that low-frequency compensation should be done before high-frequency compensation. Its N2140A product page notes that probe can be adjusted for both. Follow the probe manual, since the HF adjustment point and method vary by model.
  7. Re-check. Confirm that the top of the square wave is flat, the leading corner is square, and nothing drifts when you touch the probe lightly.

Reading the waveform

Tektronix’s probe primer gives the goal as a calibration waveform with flat tops, with no overshoot or rounding. In the manufacturers’ guides the square wave is shown in three cases: properly compensated, overcompensated and undercompensated.

What you see Likely cause What to do
Flat top, square corner Probe matched to the channel Leave the trimmer alone and take the measurement.
Spike or overshoot on the leading edge, followed by a sloping top Overcompensated Turn the trimmer back in small steps until the top is flat.
Rounded leading edge, slow rise to the top Undercompensated Turn the trimmer the other way in small steps until the corner squares up.
Flat top at low frequency but ringing or a rounded corner that the LF trimmer cannot remove High-frequency mismatch, grounding or bandwidth limit Check the ground lead, then use the HF adjustment if the probe has one.

The direction of the trimmer varies between probe designs, so do not rely on a fixed clockwise or anticlockwise rule. Watch the display and let the waveform tell you which way to go.

If the square wave will not flatten

Check the basics first

  • Attenuation switch. A probe left in 1:1 will look wrong however much you turn the trimmer.
  • Ground connection. A long or loose ground lead adds inductance and shows up as ringing on fast edges. Use the shortest ground path available.
  • Input impedance. A 10:1 passive probe is designed for a 1 MΩ channel. A 50 Ω input is the wrong load for it.
  • Probe and scope pairing. Check that the probe is intended for the scope family. A probe designed for a different input capacitance may run out of adjustment range.
  • Physical damage. A damaged tip, cable or compensation box can prevent a clean adjustment. Replace the probe rather than force it.

Know the limits of compensation

Compensation corrects the match between probe and input. It does not extend the bandwidth of the system. A probe or scope that is too slow for the signal will still round fast edges even when the comparison square wave looks perfect. For that decision, see Oscilloscope Bandwidth Explained.

Probes that ship with common bench scopes

Most general-purpose bench scopes include passive probes that need this adjustment. The table below shows what Keysight lists on its product pages for two InfiniiVision models in the IMM range. Check the current configuration before ordering, because included accessories can change between production runs.

Model Bandwidth (Keysight) Analog channels Included probes (Keysight)
Keysight DSOX1202G 70 MHz base (100 and 200 MHz need an additional option) 2 Two N2140A 200 MHz 10:1/1:1 switchable passive probes
Keysight DSOX3012A 100 MHz 2 One N2862B 150 MHz 10:1 passive probe per channel

The switchable N2140A in the first row is a good illustration of why the attenuation check matters. Keysight lists it with a 1:1 or 10:1 ratio, and with 1 MΩ input resistance at 1:1 or 10 MΩ at 10:1. Compensation only applies in the 10:1 position. The fixed 10:1 N2862B on the second model has no such switch, so the only variable there is the trimmer.

Good practice for labs and field teams

  • Make it a pre-measurement habit. A short check on the compensation terminal before each session is cheaper than repeating a measurement.
  • Label probes to channels. If a probe is permanently used on one channel, you only need to confirm, not readjust.
  • Record the check. In an accredited or audited lab, note the date and the scope and probe identifiers alongside other pre-use checks.
  • Keep the adjustment tool with the probe. A missing trimmer tool is the most common reason a probe is not adjusted at all.
  • Do not substitute for calibration. Keep the scope and probes on the calibration schedule that applies to your quality system.

For procurement teams, remember that probes are a separate item from the scope. If you are sourcing additional or higher-bandwidth probes, browse the oscilloscope probes category and match the probe’s bandwidth and input requirements to your scope. Specifications for the probes referenced here are on Keysight’s N28xxA/B passive probe datasheet, and the difference between compensation and calibration is explained in the Tektronix probe compensation article. For a broader setup walkthrough on one scope family, see Scope Keysight: Complete Guide to Calibration and Setup.

Frequently asked questions

How often should an oscilloscope probe be compensated?

Compensate a probe when it is new, when it is moved to a different channel or scope, and before measurements where edge shape matters. Tektronix also recommends it after significant environmental changes or probe maintenance. A quick check on the compensation terminal before each session is a sensible habit.

Why must the probe be set to 10:1 before compensating?

The compensation network belongs to the 10:1 divider. On a switchable probe such as the Keysight N2140A, the 1:1 position bypasses that divider, so adjusting the trimmer there gives a misleading result. Set 10:1, then adjust, and match the channel's attenuation setting.

What square wave frequency is used for compensation?

Keysight's datasheet for its N28xxA/B passive probes specifies a 1 kHz square wave source for low-frequency compensation, with two to three cycles displayed. Scopes with a front-panel compensation terminal supply a suitable signal. Check your scope manual for the exact output.

Does compensating a probe replace probe calibration?

No. Tektronix describes compensation as an adjustment that matches probe capacitance to the scope input, while calibration compares measurements to a known standard. Compensation does not change calibration status, so a probe still needs its own calibration schedule where your quality system requires one.

The square wave will not flatten. What should I check?

Check that the probe is in the 10:1 position, the channel is set to 1 MΩ input, the ground lead is short and secure, and the probe is intended for that scope. If the HF adjustment is available, use it after the LF adjustment. A damaged probe should be replaced.