{"product_id":"keysight-8495b-002-step-attenuator-dc-18-ghz-70-db-sma","title":"Keysight 8495B\/002 Step Attenuator, DC-18 GHz, 70 dB, SMA","description":"\u003ch2\u003eKeysight 8495B\/002 Step Attenuator, DC-18 GHz, 70 dB, SMA\u003c\/h2\u003e\n\u003ch3\u003eOverview\u003c\/h3\u003e\n\u003cp\u003eRepeatability is ±0.03 dB maximum over 5 million cycles per section, keeping measurement uncertainty low in test systems. Insertion loss at 0 dB is 1.66 dB and maximum SWR is 1.70 up to 18 GHz for this model.\u003c\/p\u003e\n\u003cp\u003eThe attenuator handles 1 W average and 100 W peak RF input power, and it can be cascaded with a second attenuator or the 11716C SMA interconnect kit for greater dynamic range with finer steps. Option UK6 adds NIST traceable SWR and attenuation data.\u003c\/p\u003e\n\u003ch3\u003eFeatures\u003c\/h3\u003e\n\u003cul\u003e\n\u003cli\u003eDC to 18 GHz frequency range\u003c\/li\u003e\n\u003cli\u003e0 to 70 dB in 10 dB steps\u003c\/li\u003e\n\u003cli\u003eSMA female connectors (option 002)\u003c\/li\u003e\n\u003cli\u003eRepeatability ±0.03 dB max, 5 million cycles per section\u003c\/li\u003e\n\u003cli\u003eInsertion loss 1.66 dB at 0 dB\u003c\/li\u003e\n\u003cli\u003eMaximum SWR 1.70\u003c\/li\u003e\n\u003cli\u003e1 W average, 100 W peak input power\u003c\/li\u003e\n\u003cli\u003eCascadable with 11716C interconnect kit\u003c\/li\u003e\n\u003c\/ul\u003e\n\u003ch3\u003eApplications\u003c\/h3\u003e\n\u003cul\u003e\n\u003cli\u003eRF and microwave signal level control\u003c\/li\u003e\n\u003cli\u003eReceiver and link budget testing\u003c\/li\u003e\n\u003cli\u003eAttenuation in automated test setups\u003c\/li\u003e\n\u003c\/ul\u003e\n\u003ch3\u003eKeysight 8495B\/002 Step Attenuator, DC-18 GHz, 70 dB, SMA Technical Specifications\u003c\/h3\u003e\n\u003ctable\u003e\n\u003ctbody\u003e\n\u003ctr\u003e\n\u003cth\u003eFrequency range\u003c\/th\u003e\n\u003ctd\u003eDC to 18 GHz\u003c\/td\u003e\n\u003c\/tr\u003e\n\u003ctr\u003e\n\u003cth\u003eAttenuation range\u003c\/th\u003e\n\u003ctd\u003e0 to 70 dB\u003c\/td\u003e\n\u003c\/tr\u003e\n\u003ctr\u003e\n\u003cth\u003eAttenuation step\u003c\/th\u003e\n\u003ctd\u003e10 dB\u003c\/td\u003e\n\u003c\/tr\u003e\n\u003ctr\u003e\n\u003cth\u003eConnector option\u003c\/th\u003e\n\u003ctd\u003e002 = SMA female\u003c\/td\u003e\n\u003c\/tr\u003e\n\u003ctr\u003e\n\u003cth\u003eInsertion loss at 0 dB\u003c\/th\u003e\n\u003ctd\u003e1.66 dB\u003c\/td\u003e\n\u003c\/tr\u003e\n\u003ctr\u003e\n\u003cth\u003eMaximum SWR\u003c\/th\u003e\n\u003ctd\u003e1.70\u003c\/td\u003e\n\u003c\/tr\u003e\n\u003ctr\u003e\n\u003cth\u003eMaximum input power\u003c\/th\u003e\n\u003ctd\u003e1 W average, 100 W peak\u003c\/td\u003e\n\u003c\/tr\u003e\n\u003ctr\u003e\n\u003cth\u003eRepeatability\u003c\/th\u003e\n\u003ctd\u003e±0.03 dB max (5 million cycles per section)\u003c\/td\u003e\n\u003c\/tr\u003e\n\u003ctr\u003e\n\u003cth\u003eOperating life\u003c\/th\u003e\n\u003ctd\u003e5 million cycles per section\u003c\/td\u003e\n\u003c\/tr\u003e\n\u003ctr\u003e\n\u003cth\u003eSeries connection\u003c\/th\u003e\n\u003ctd\u003eKeysight 11716C SMA interconnect kit\u003c\/td\u003e\n\u003c\/tr\u003e\n\u003c\/tbody\u003e\n\u003c\/table\u003e\n\n\u003cp\u003eManufacturer information: \u003ca href=\"https:\/\/www.keysight.com\/gb\/en\/product\/8495B\/manual-attenuator-18-ghz-70-db-10-db-steps.html\" target=\"_blank\" rel=\"noopener\"\u003e8495B on keysight.com\u003c\/a\u003e.\u003c\/p\u003e","brand":"Keysight","offers":[{"title":"Default Title","offer_id":67776456720571,"sku":"IMM_22449","price":1058.0,"currency_code":"USD","in_stock":true}],"thumbnail_url":"\/\/cdn.shopify.com\/s\/files\/1\/0777\/2635\/1547\/files\/keysight-8495b-002-manual-attenuator-image-600x600-1.jpg?v=1792553493","url":"https:\/\/indomultimeter.net\/products\/keysight-8495b-002-step-attenuator-dc-18-ghz-70-db-sma","provider":"Indo Multimeter","version":"1.0","type":"link"}