{"product_id":"fluke-pressurements-p3112-kgcm2-deadweight-tester-4-to-140-kgf-cm2","title":"Fluke Pressurements P3112-KGCM2 Deadweight Tester, 4 to 140 kgf\/cm2","description":"\u003ch2\u003eFluke Pressurements P3112-KGCM2 Deadweight Tester\u003c\/h2\u003e\u003ch3\u003eOverview\u003c\/h3\u003e\n\u003cp\u003eThe Fluke Pressurements P3112-KGCM2 is a single-PCU hydraulic deadweight tester in the P3100 series, an instrument family built on more than five decades of primary pressure standard design. The tester works on the piston-gauge principle: a precision-lapped piston and cylinder assembly supports calibrated weight masses, and when the piston spins and floats freely the upward force from system pressure balances the applied weights. Pressure is then calculated from the total supported mass divided by the effective piston area.\u003c\/p\u003e\n\u003cp\u003eAccuracy is 0.015% of reading, with an optional 0.008% of reading grade, and certified calibrations are traceable to international standards laboratories such as NIST. The piston is made of tungsten carbide with a nickel binder running in a hardened martensitic steel cylinder, a combination chosen for wear resistance and long-term stability. Each instrument is trimmed to local gravity when ordered, and a built-in priming pump and screw press handle larger-volume devices under test.\u003c\/p\u003e\n\u003ch3\u003eFeatures\u003c\/h3\u003e\n\u003cul\u003e\n\u003cli\u003e4 to 140 kgf\/cm2 pressure range, oil operated\u003c\/li\u003e\n\u003cli\u003eAccuracy of 0.015% of reading (0.008% optional)\u003c\/li\u003e\n\u003cli\u003eSingle piston-cylinder unit (PCU) configuration\u003c\/li\u003e\n\u003cli\u003eTungsten carbide piston with nickel binder, 15 g\/cm3 density\u003c\/li\u003e\n\u003cli\u003eSeries 3 non-magnetic stainless steel standard masses\u003c\/li\u003e\n\u003cli\u003eBuilt-in hand pump and screw press for larger volumes\u003c\/li\u003e\n\u003cli\u003eMounted spirit level and adjustable feet\u003c\/li\u003e\n\u003cli\u003eUnits trimmed to local gravity at no extra cost\u003c\/li\u003e\n\u003c\/ul\u003e\n\u003ch3\u003eApplications\u003c\/h3\u003e\n\u003cul\u003e\n\u003cli\u003ePrimary-standard calibration of pressure transducers, transmitters and gauges\u003c\/li\u003e\n\u003cli\u003eCalibration of pressure switches in calibration laboratories\u003c\/li\u003e\n\u003cli\u003eReference pressure generation for industrial instrument shops\u003c\/li\u003e\n\u003c\/ul\u003e\n\u003ch3\u003eFluke Pressurements P3112-KGCM2 Deadweight Tester Technical Specifications\u003c\/h3\u003e\n\u003ctable\u003e\u003ctbody\u003e\n\u003ctr\u003e\n\u003ctd\u003ePressure range\u003c\/td\u003e\n\u003ctd\u003e4 to 140 kgf\/cm2\u003c\/td\u003e\n\u003c\/tr\u003e\n\u003ctr\u003e\n\u003ctd\u003ePCU configuration\u003c\/td\u003e\n\u003ctd\u003eSingle piston-cylinder unit\u003c\/td\u003e\n\u003c\/tr\u003e\n\u003ctr\u003e\n\u003ctd\u003eOperating fluid\u003c\/td\u003e\n\u003ctd\u003eMineral oil\u003c\/td\u003e\n\u003c\/tr\u003e\n\u003ctr\u003e\n\u003ctd\u003eAccuracy\u003c\/td\u003e\n\u003ctd\u003e±0.015% of reading (±0.008% optional)\u003c\/td\u003e\n\u003c\/tr\u003e\n\u003ctr\u003e\n\u003ctd\u003ePiston material\u003c\/td\u003e\n\u003ctd\u003eTungsten carbide with nickel binder, 15 g\/cm3\u003c\/td\u003e\n\u003c\/tr\u003e\n\u003ctr\u003e\n\u003ctd\u003eCylinder material\u003c\/td\u003e\n\u003ctd\u003eHardened martensitic steel\u003c\/td\u003e\n\u003c\/tr\u003e\n\u003ctr\u003e\n\u003ctd\u003eStandard weight material\u003c\/td\u003e\n\u003ctd\u003eSeries 3 non-magnetic austenitic stainless steel, 7.8 g\/cm3\u003c\/td\u003e\n\u003c\/tr\u003e\n\u003ctr\u003e\n\u003ctd\u003ePressure range increment\u003c\/td\u003e\n\u003ctd\u003e0.1 kgf\/cm2\u003c\/td\u003e\n\u003c\/tr\u003e\n\u003ctr\u003e\n\u003ctd\u003eReservoir volume\u003c\/td\u003e\n\u003ctd\u003e150 cm3 (9.2 in3)\u003c\/td\u003e\n\u003c\/tr\u003e\n\u003ctr\u003e\n\u003ctd\u003eScrew press displacement\u003c\/td\u003e\n\u003ctd\u003e5.5 cc\u003c\/td\u003e\n\u003c\/tr\u003e\n\u003ctr\u003e\n\u003ctd\u003ePump displacement\u003c\/td\u003e\n\u003ctd\u003e4.7 cc per stroke\u003c\/td\u003e\n\u003c\/tr\u003e\n\u003ctr\u003e\n\u003ctd\u003eTest port adapters\u003c\/td\u003e\n\u003ctd\u003e1\/8, 1\/4, 3\/8 and 1\/2 NPT or BSP\u003c\/td\u003e\n\u003c\/tr\u003e\n\u003ctr\u003e\n\u003ctd\u003eThermal expansion, oil piston\/cylinder\u003c\/td\u003e\n\u003ctd\u003e16.5 ppm\/°C\u003c\/td\u003e\n\u003c\/tr\u003e\n\u003ctr\u003e\n\u003ctd\u003eGravity setting\u003c\/td\u003e\n\u003ctd\u003eTrimmed to local gravity; 980.665 cm\/s2 standard if unspecified\u003c\/td\u003e\n\u003c\/tr\u003e\n\u003c\/tbody\u003e\u003c\/table\u003e\n\n\u003cp\u003eManufacturer information: \u003ca href=\"https:\/\/www.fluke.com\/en-us\/product\/calibration-tools\/pressure-calibrators\/fluke-calibration-p3100\" target=\"_blank\" rel=\"noopener\"\u003eFluke Calibration P3100 Hydraulic Deadweight Testers — fluke.com\u003c\/a\u003e\u003c\/p\u003e","brand":"Fluke","offers":[{"title":"Default Title","offer_id":67776330825915,"sku":"IMM_37477764","price":6060.0,"currency_code":"USD","in_stock":true}],"thumbnail_url":"\/\/cdn.shopify.com\/s\/files\/1\/0777\/2635\/1547\/files\/fluke-pressurements-3-1_c05cb7bd-0ed4-40f4-b558-5783fe4d8c69.jpg?v=1792552835","url":"https:\/\/indomultimeter.net\/products\/fluke-pressurements-p3112-kgcm2-deadweight-tester-4-to-140-kgf-cm2","provider":"Indo Multimeter","version":"1.0","type":"link"}