{"product_id":"fluke-calibration-5618b-9-p-rtd-probe-9-in-x-3-2-mm-200-to-500-c","title":"Fluke Calibration 5618B-9-P RTD Probe, 9 in x 3.2 mm, -200 to 500 °C","description":"\u003ch2\u003eFluke Calibration 5618B-9-P RTD Probe\u003c\/h2\u003e\u003ch3\u003eOverview\u003c\/h3\u003e\n\u003cp\u003eThe Fluke Calibration 5618B-9-P RTD Probe uses a nominal 100 Ω reference-grade element with a 0.003923 Ω\/Ω\/°C temperature coefficient. Calibrated accuracy (k=2) is ±0.066 °C at 0 °C, ±0.026 °C at -200 °C and ±0.219 °C at 500 °C, with drift of ±0.1 °C when used periodically to 500 °C.\u003c\/p\u003e\n\u003cp\u003eEach probe arrives with a manufacturer's accredited calibration including ITS-90 coefficients and R vs T values in 1 °C increments, ready for Fluke Calibration 1523, 1524, 1502A, 1529 or 1560 readouts. Hysteresis is below 0.01 °C at 0 °C, the 22 AWG PTFE leads run six feet, and the P termination is an INFO-CON connector for 1523\/1524 thermometers. These probes are also used to calibrate 9132 and 9133 infrared calibrator targets.\u003c\/p\u003e\n\u003ch3\u003eFeatures\u003c\/h3\u003e\n\u003cul\u003e\n\u003cli\u003eSmall diameter 3.2 mm (1\/8 in) sheath for faster response\u003c\/li\u003e\n\u003cli\u003e-200 °C to 500 °C temperature range\u003c\/li\u003e\n\u003cli\u003eCalibrated accuracy ±0.066 °C at 0 °C (k=2)\u003c\/li\u003e\n\u003cli\u003eReference-grade element on the 0.003923 Ω\/Ω\/°C curve\u003c\/li\u003e\n\u003cli\u003eHysteresis below 0.01 °C at 0 °C\u003c\/li\u003e\n\u003cli\u003eINFO-CON (P) termination for Fluke Calibration 1523\/1524\u003c\/li\u003e\n\u003cli\u003eAccredited calibration with ITS-90 coefficients included\u003c\/li\u003e\n\u003cli\u003eAlso suited to calibrating 9132 and 9133 IR calibrator targets\u003c\/li\u003e\n\u003c\/ul\u003e\n\u003ch3\u003eApplications\u003c\/h3\u003e\n\u003cul\u003e\n\u003cli\u003eSecondary-level reference for industrial temperature calibration\u003c\/li\u003e\n\u003cli\u003eCalibrating IR calibrator targets such as the 9132 and 9133\u003c\/li\u003e\n\u003cli\u003eUse with Fluke Calibration thermometer readouts in labs and plants\u003c\/li\u003e\n\u003c\/ul\u003e\n\u003ch3\u003eFluke Calibration 5618B-9-P RTD Probe Technical Specifications\u003c\/h3\u003e\n\u003ctable\u003e\n\u003ctbody\u003e\n\u003ctr\u003e\n\u003cth\u003eModel\u003c\/th\u003e\n\u003ctd\u003e5618B-9-P, small diameter industrial RTD\u003c\/td\u003e\n\u003c\/tr\u003e\n\u003ctr\u003e\n\u003cth\u003eTemperature range\u003c\/th\u003e\n\u003ctd\u003e-200 °C to 500 °C\u003c\/td\u003e\n\u003c\/tr\u003e\n\u003ctr\u003e\n\u003cth\u003eNominal resistance\u003c\/th\u003e\n\u003ctd\u003e100 Ω at 0 °C\u003c\/td\u003e\n\u003c\/tr\u003e\n\u003ctr\u003e\n\u003cth\u003eTemperature coefficient\u003c\/th\u003e\n\u003ctd\u003e0.003923 Ω\/Ω\/°C nominal\u003c\/td\u003e\n\u003c\/tr\u003e\n\u003ctr\u003e\n\u003cth\u003eCalibrated accuracy (k=2)\u003c\/th\u003e\n\u003ctd\u003e+\/-0.026 °C at -200 °C; +\/-0.066 °C at 0 °C; +\/-0.219 °C at 500 °C\u003c\/td\u003e\n\u003c\/tr\u003e\n\u003ctr\u003e\n\u003cth\u003eDrift\u003c\/th\u003e\n\u003ctd\u003e±0.1 °C when used periodically to 500 °C\u003c\/td\u003e\n\u003c\/tr\u003e\n\u003ctr\u003e\n\u003cth\u003eHysteresis\u003c\/th\u003e\n\u003ctd\u003e\u0026lt;0.01 °C at 0 °C\u003c\/td\u003e\n\u003c\/tr\u003e\n\u003ctr\u003e\n\u003cth\u003eSheath\u003c\/th\u003e\n\u003ctd\u003e229 mm x 3.2 mm (9 in x 1\/8 in), 316 stainless steel\u003c\/td\u003e\n\u003c\/tr\u003e\n\u003ctr\u003e\n\u003cth\u003eLeads\u003c\/th\u003e\n\u003ctd\u003e22 AWG PTFE, 6 ft\u003c\/td\u003e\n\u003c\/tr\u003e\n\u003ctr\u003e\n\u003cth\u003eTermination\u003c\/th\u003e\n\u003ctd\u003eP - INFO-CON connector for 1523\/1524\u003c\/td\u003e\n\u003c\/tr\u003e\n\u003ctr\u003e\n\u003cth\u003eTime constant\u003c\/th\u003e\n\u003ctd\u003e4 s maximum (63.2%, water at 3 fps)\u003c\/td\u003e\n\u003c\/tr\u003e\n\u003ctr\u003e\n\u003cth\u003eCalibration\u003c\/th\u003e\n\u003ctd\u003eManufacturer's accredited calibration, ITS-90 coefficients\u003c\/td\u003e\n\u003c\/tr\u003e\n\u003c\/tbody\u003e\n\u003c\/table\u003e\n\n\u003cp\u003eManufacturer information: \u003ca href=\"https:\/\/www.fluke.com\/en\/product\/accessories\/temperature\/5618b\" target=\"_blank\" rel=\"noopener\"\u003e5618B on fluke.com\u003c\/a\u003e.\u003c\/p\u003e","brand":"Fluke Calibration","offers":[{"title":"Default Title","offer_id":67776911114427,"sku":"IMM_73758369","price":772.0,"currency_code":"USD","in_stock":true}],"thumbnail_url":"\/\/cdn.shopify.com\/s\/files\/1\/0777\/2635\/1547\/files\/fluke-calibration-5618b-9-p.jpg?v=1792555558","url":"https:\/\/indomultimeter.net\/products\/fluke-calibration-5618b-9-p-rtd-probe-9-in-x-3-2-mm-200-to-500-c","provider":"Indo Multimeter","version":"1.0","type":"link"}