Key specs from the product description

Model
ADT875-1210-FS-110V
Temperature range
100°C to 1210°C
Stability
±0.1°C
Display accuracy
±1.2°C (100 to 600°C); ±1.6°C (900°C); ±2.0°C (1210°C)
Axial uniformity (20 mm zone)
±0.6°C at 100°C to ±1.5°C at 1210°C
Power
90-254 VAC, 45-65 Hz, 580 W
Full description ↓
Additel ADT875-1210-FS-110V Thermocouple Calibration Furnace Overview The Additel ADT875-1210-FS-110V is a thermocouple calibration furnace covering 100°C to 1210°C with a stability of ±0.1°C. It uses three-zone temperature control...

Need a formal quote or a volume price?

For purchase orders, proforma invoices or several units. We typically reply within one working day.

Available to orderWe ship worldwide Typical response within one working day. The quote confirms the model, price and availability.
Add company, phone or a note
  • Typical reply: one working day
  • Proforma invoice on request
  • Pay by card or bank transfer
  • 30-day money-back on defects
We ship worldwide · Prices in USD
  1. 1Send the requestTakes a minute
  2. 2Get your priceTypically one working day
  3. 3Pay, we shipCard or bank transfer
  • 30-day money-back if the item arrives defective, damaged or wrong. Policy
  • Traceable calibration documentation ships with instruments
  • Pay by card or bank transfer, invoice on request
  • We ship worldwide. Destination confirmed with your quote

Datasheets & manuals

Need the datasheet? Add the request to your quote. Ask for the datasheet
ShippingWe ship worldwide
DocumentsTraceable calibration documentation
Quote responseTypically within one working day

Additel ADT875-1210-FS-110V Thermocouple Calibration Furnace

Overview

The Additel ADT875-1210-FS-110V is a thermocouple calibration furnace covering 100°C to 1210°C with a stability of ±0.1°C.

It uses three-zone temperature control with Advanced Adaptive Control, and is offered with an optional Process Calibrator (PC).

Features

  • Temperature range 100°C to 1210°C
  • Stability ±0.1°C
  • Three-zone control with Advanced Adaptive Control (wind tunnel control combined with 3-zone temperature control)
  • Process Calibrator (PC) option: one reference TC channel and up to four channels under test; channels 1 and 2 also measure mA and voltage, perform switch tests and source 24V DC
  • Self-calibration with the PC option
  • HART communication optional on PC models
  • 110V version (ordering options: 110V or 220V)

Applications

Calibration of thermocouples in process control, laboratory testing and industrial maintenance.

Additel ADT875-1210-FS-110V Thermocouple Calibration Furnace Technical Specifications

Model ADT875-1210-FS-110V
Temperature range 100°C to 1210°C
Stability ±0.1°C
Display accuracy ±1.2°C (100 to 600°C); ±1.6°C (900°C); ±2.0°C (1210°C)
Axial uniformity (20 mm zone) ±0.6°C at 100°C to ±1.5°C at 1210°C
Power 90-254 VAC, 45-65 Hz, 580 W

Price $6,540.00 USD as listed on 27 Nov 2026. Request a firm quote: https://indomultimeter.net/products/additel-adt875-1210-fs-110v-thermocouple-calibration-furnace-100-1210c

Before you order

Is the price final?

The price shown is our current price in USD. Your quote confirms price and availability for your quantity.

How is my item supplied?

Items are ordered from the manufacturer for you. The quote confirms availability. Typical response is within one working day.

Do you provide calibration certificates?

Instruments ship with calibration documentation traceable to national standards. Ask in your request if you need a specific certificate.

How can I pay?

By card or by bank transfer. We can send a proforma invoice for your purchasing department.

Do you ship to my country?

We ship worldwide. Put your destination country in your enquiry and we confirm delivery with your quote.

What if it arrives defective, damaged or wrong?

You get a 30-day money-back guarantee for items that arrive defective, damaged or wrong. See the refund policy.