How to Run a Pressure Hold Test with a Reference Recorder

A pressure hold test is passed when the corrected pressure stays within the allowed decay over the hold period, not when the raw gauge reading looks flat. Log pressure and temperature at a fixed rate, let the system stabilise thermally, then correct the end reading for temperature (and for barometric change if you read gauge pressure) before you call it a leak.
What a hold test actually measures
A hold test pressurises a closed volume, isolates it from the source, and watches the pressure for a defined time. Any decay is then compared with an acceptance limit. That limit comes from your procedure, client specification or governing code. This article does not set one, and you should not borrow a limit from another system.
The difficulty is that a sealed volume does not only lose pressure through leaks. Its pressure also moves with temperature, and a gauge reading moves with the atmosphere. A hold test that ignores both will fail good systems on cold nights and pass leaking ones on warm afternoons.
Equipment and setup
- A pressure reference with a range close to the test pressure. A module sized for the test pressure resolves small decay better than one with a range ten times larger.
- A logger, not a human. Manual readings every few minutes miss the shape of the decay. A recorder shows whether the pressure is settling, falling steadily or stepping down.
- A temperature record. Pressure alone cannot separate a leak from thermal drift. Log temperature alongside it, and know where that temperature is measured.
- A barometric reference if you read gauge pressure and the hold is long.
- Isolation valves you trust. A weeping isolation valve looks exactly like a leak in the system under test. Where practical, use a double block or a blind.
The AMETEK Crystal Engineering nVision reference recorder is one example of this class of instrument. AMETEK states that it is temperature compensated with accuracy not degraded from -20 to 50 °C, records temperature as well as pressure, stores up to 1 million data points at up to 10 readings per second, and exports data to a spreadsheet. See the manufacturer’s hydrostatic test note for its description of this use.
Step-by-step procedure
- Check the reference first. Confirm the recorder is within its calibration interval and zero it at atmospheric pressure with the port open, if the module is a gauge type.
- Connect the reference as close to the test volume as possible. Long, small-bore tubing adds lag and can hide a pressure gradient. Tighten fittings per the manufacturer’s instructions and leak-check the connection itself with the reference alone.
- Set the logging interval. Choose a rate fast enough to see the pressurisation transient and slow enough to keep the file manageable. A common approach is a fast interval during fill, then a slower one for the hold. Check your recorder allows this, or use one rate that suits both.
- Pressurise slowly to the test pressure. Rapid filling with gas heats it through compression. A hydraulic fill needs air bled out first, because trapped air makes the pressure drift for a long time.
- Isolate the source and start the hold clock only after stabilisation. Wait until the pressure and temperature traces have both flattened. Compression heat decays within minutes for small volumes and much longer for large pipework.
- Record the start. Note pressure, temperature, barometric pressure and the time on the test sheet, as well as in the log.
- Hold for the specified time. Do not touch valves and keep people away from the line. Avoid direct sun or heat sources on the piping during the hold.
- Record the end values. Take the same four readings. Save the raw log file before any processing.
- Correct, then judge. Apply the corrections below and compare the corrected decay with the acceptance limit.
- Depressurise safely through the vent route in your procedure, and note it on the record.
Correcting for temperature on a gas test
For a fixed volume of gas, absolute pressure is proportional to absolute temperature. Convert gauge pressure to absolute by adding atmospheric pressure, convert °C to kelvin by adding 273.15, and compare the end reading with what temperature alone would predict.
Expected end pressure (absolute) = start pressure (absolute) × end temperature (K) ÷ start temperature (K).
A worked example, using 14.7 psi as atmospheric pressure for simplicity:
| Item | Start of hold | End of hold (30 min) |
|---|---|---|
| Gauge pressure | 100.0 psig | 99.6 psig |
| Absolute pressure | 114.7 psia | 114.3 psia |
| Gas temperature | 20.0 °C (293.15 K) | 19.0 °C (292.15 K) |
| Expected from cooling alone | n/a | 114.7 × 292.15 ÷ 293.15 = 114.31 psia |
| Residual (measured minus expected) | n/a | about -0.01 psi |
The raw reading fell 0.4 psi, which looks like a failure. But a 1 K drop in a 114.7 psia gas volume accounts for about 0.39 psi, so the corrected residual is close to zero. This system held. If the residual had been -0.3 psi, the leak would be real even though the raw drop looks the same.
The correction is only as good as the temperature you use. A sensor on the instrument body may not follow the gas or the pipe wall. Where the line is large, outdoors or in sun, use a temperature measured on the line, and say in the report where it was taken.
Correcting for barometric change
If you read gauge pressure, the reading is absolute pressure minus the atmosphere. A sealed system holds its absolute pressure, so when the atmosphere rises the gauge reading falls, and the reverse. A 10 mbar change in barometric pressure is about 0.145 psi. Over a multi-hour test on a day with a moving weather front, that is enough to matter on low-pressure tests.
Two practical fixes exist. Either read absolute pressure, or log barometric pressure alongside gauge pressure and correct the data. Several nVision configurations in the catalogue carry a BARO module in the part number, which is the second approach. Check the exact configuration before you order, because gauge-only recorders do not read atmospheric pressure.
Common error sources
| Source | What you see | How to handle it |
|---|---|---|
| Compression heating | Fast initial decay that slows to a curve | Wait for temperature to settle before starting the hold |
| Ambient temperature swing | Pressure rises and falls with the day | Log temperature and correct; shield the line from sun |
| Barometric change | Slow drift in gauge reading | Read absolute or log barometric pressure |
| Leaking isolation valve | Steady decay that looks like a leak | Block and bleed, blind, or test the valve separately |
| Fitting leak at the reference | Decay present even on a known-good system | Leak-check the connection with the reference alone first |
| Trapped air in a liquid test | Slow, irregular drift | Bleed air at high points before pressurising |
| Oversized module | Coarse steps in the pressure trace | Use a module whose range suits the test pressure |
Liquid (hydrostatic) tests behave differently
A liquid is far less compressible than a gas, so a trapped liquid reacts strongly to small temperature changes. A fraction of a degree can move the pressure noticeably, in either direction. That is why hydrostatic tests are often documented with temperature as well as pressure, and why a rising reading during a hold is just as important to explain as a falling one. Do not apply the ideal-gas correction above to a liquid test. Follow the method in your governing procedure, and use the temperature trace to explain the shape of the curve.
Reading the log
Plot pressure and temperature on the same time axis before doing any arithmetic. Three patterns cover most results.
- Pressure tracks temperature, with no step. The decay is thermal. Correct it and it should vanish.
- Straight-line decay while temperature is flat. This is a leak. The slope gives you a rate you can report.
- A sudden step. Something moved: a valve, a relief device, a gasket seating, or someone adjusting the line. Find the cause before accepting the test.
Keep the raw log with the test record. A data file with a visible trend is stronger evidence than two hand-written numbers. AMETEK describes its recorder as storing data in non-volatile memory, so data survives battery depletion during a test; see its application note on replacing a chart recorder.
Choosing a recorder for this work
Match the module range to the test pressure, decide whether you need a barometric module or an RTD input for temperature, and confirm the calibration certificate and traceability you need for your quality system. For a dual-range unit that covers a low-pressure leak check and a higher-pressure strength test, the Crystal Engineering NV-4AA-30PSI-1KPSI-B reference recorder carries 30 psi and 1000 psi modules. For a single-range test with atmospheric correction, the NV-4AA-300PSI-BARO-B adds a barometric module, and the NV-4AA-RTD100-30PSI-BARO-B combines a 30 psi module with an RTD input and a barometric module. Browse the full range under multifunction dataloggers.
If your test also involves checking the gauges on the system, our guide to pressure gauge calibration covers that procedure. For sizing the reference itself, see how to choose a pressure calibrator.
Frequently asked questions
How long should a pressure hold test last?
The duration is set by your test procedure, client specification or governing code, not by the instrument. Longer holds resolve smaller leaks but pick up more temperature and barometric drift. Whatever time you use, start the clock only after pressure and temperature have stabilised.
Why does pressure drop during a hold test when there is no leak?
In a sealed gas system, pressure follows absolute temperature. Compression heat from filling, then cooling, lowers pressure with no leak. A gauge reading also falls when the atmosphere rises. Logging temperature and barometric pressure lets you correct for both and see the true decay.
Can I use the gas temperature correction for a liquid test?
No. The ideal-gas relationship applies to gases. A trapped liquid is far less compressible, so small temperature changes move pressure strongly. For hydrostatic tests, follow the method in your governing procedure and use the logged temperature trace to explain pressure movements.
Do I need a barometric module for a hold test?
Not always. If you read absolute pressure, or the hold is short and the pressure is high, barometric change may be negligible. On low-pressure tests over several hours, a 10 mbar atmospheric change is about 0.145 psi, which can matter, so a barometric reference is worth having.